11 respiratory targets and 1,405 specimens. What the German clinical study means for JOYTEST
This clinical study at three German sites compared a broad antigen panel from our manufacturing partner DeepBlue with reference RT-PCR. In the main set of 1,405 specimens, the panel achieved 95–98% sensitivity across 11 viral and bacterial targets.
Download the clinical study summary for physicians (PDF)
Study sites and methods
Dr Haidong Teng’s practice in Mühlheim am Main and the Klinikum Hanau laboratory handled sampling and antigen testing. MVZ Labor Dr. Limbach & Kollegen in Heidelberg performed reference RT-PCR with Seegene Allplex RV Master Assay and Allplex Respiratory Panel 4. The protocol specifies a nasal swab for antigen testing and a paired nasopharyngeal swab for RT-PCR. The manufacturer’s report cites Annex XIII to Regulation (EU) 2017/746 (IVDR) and ISO 20916:2019 good study practice. [1]
Of 1,575 specimens evaluated, the main set included 1,100 PCR-positive specimens (100 per target) and 305 specimens negative for all 11 targets. PCR-positive specimens were collected within seven days of symptom onset. [1]
11-in-1 panel results vs RT-PCR
| Target | Sensitivity | 95% confidence interval |
|---|---|---|
| SARS-CoV-2 | 97% | 91.55–98.97% |
| Influenza A | 96% | 90.16–98.43% |
| Influenza B | 95% | 88.82–97.85% |
| Respiratory syncytial virus (RSV) | 97% | 91.55–98.97% |
| Adenovirus | 98% | 93.00–99.45% |
| Human metapneumovirus (hMPV) | 96% | 90.16–98.43% |
| Parainfluenza virus (PIV) | 96% | 90.16–98.43% |
| Mycoplasma pneumoniae | 97% | 91.55–98.97% |
| Rhinovirus | 95% | 88.82–97.85% |
| Streptococcus pneumoniae | 95% | 88.82–97.85% |
| Legionella pneumophila | 96% | 90.16–98.43% |
Observed specificity was 100% per target (0 false positives among 1,305 RT-PCR-negative specimens per target, 95% CI 99.71–100%). Depending on the target, the antigen test missed 2–5 of 100 PCR-positive specimens. [1]
What the study means for JOYTEST respiratory panels
DeepBlue is our direct manufacturing partner. With a shared manufacturer and overlapping detection targets, the study is relevant to most professional JOYTEST respiratory panels. The values in the table apply to the 11-in-1 panel evaluated in the study. Each JOYTEST model has its own performance data in its instructions for use.
Why the study did not include pertussis
The German study evaluated the 11-in-1 panel and began before JOYMED added JOYTEST 12-in-1 to its range. The 12-in-1 panel covers all 11 targets in the study and adds detection of Bordetella pertussis, the bacterium that causes whooping cough. The instructions for use for JOYTEST 12-in-1 report 98.23% positive agreement (111/113) and 235/235 negative agreement with the comparator kit for this target. [2]
JOYMED prepared this summary with the manufacturer’s written permission. The full original report is available from JOYMED on request.
References
- Anhui DeepBlue Medical Technology Co., Ltd. Eleven Respiratory Pathogens Antigen Test Kit (LFA) Clinical Performance Study Report. CPSP 250821-09, v. 1.0, 2026-06-17. pp. 6–8, 17, 20, 22–27, 32.
- JOYTEST 12-in-1 instructions for use, Czech edition (JOY121P1), IFU-RP12Ag-NNP-01, A/0, 2025-08-18. Section 4.9.